•Rapid
and specific typing of Verotoxin I (VT1) and II (VT2) produced by
Verotoxin producing Escherichia coli
Loopamp® Verotoxin Typing
Kit uses Verotoxin (VT1 and VT2, also known as Shiga toxins Stx1 and
Stx2, or Shiga-like toxins SLT1 and SLT2) genes as the target sequences.
By using four primers to recognize six distinct regions in each gene,
these tests are highly specific. Due to the high amplification
efficiency of LAMP, determination of Verotoxin type is completed within
one hour after a 10 minute DNA isolation from microbial culture or
colonies.
•Amplification and
detection in one single-tube with real-time results.
With Loopamp® Realtime
Turbidimeters, amplification and real-time detection are achieved in
one step.
Since electrophoresis is not required, the entire process for each type is
completed within one reaction tube, which greatly reduces the possibility of
cross-contamination frequently associated with nucleic acid
amplification technologies.
•Alternative
method of detection
Because LAMP is an isothermal reaction, the testing can
also be performed in a water bath or a heat block set at 65°C.
The endpoint results can be visualized by adding Loopamp® Fluorescent
Detection Reagent (sold separately) to the initial master mix.
• For more detailed information, view
the Package Insert
|